Three different
test
methods which require no
special instrumentation are described below.
Test Method I
1)
Place the collimator in automatic mode.
2)
Center a small cassette, 18 x 24 cm
(8 x 10"), in the cassette tray.
4)
3) Center a larger cassette on the table top.
5)
Make an exposure.
Process the film.
If the x-ray field
size exceeds the film size in the cassette
tray, use triangulation and the radiograph
from the table top to determine the actual
x-ray field size at the cassette tray.
6)
Repeat steps 2-5 with the cassettes rotated
90 degrees in the cassette tray and on the
table top.
7)
Repeat steps 2-6 with a larger cassette in
the cassette tray.
Test Method II
1)
Place the collimator in automatic mode.
2)
Center a small cassette 18 x 24 cm (8 x 10")
in the cassette tray.
3) Disable the automatic PBL feature.
4) Remove the small cassette and place a
larger cassette with film in the cassette
tray.
5) Make an exposure and measure the x-ray
field size on the film.
6) Repeat steps l-5 with the two cassettes
rotated 90 degrees in the cassette tray.
Repeat steps
1
-6 with a larger cassette'
7) 30 x 35 cm (11 x 14") in cassette tray
initially.
Test Method III
Since the Numerical Indicator accuracy was verified
in I-6-A-C, these indicators can be used to verify PBL
sizing.
2)
1) Place the collimator in automatic mode.
Center a small cassette, 18 x 24 cm (8 x 10"))
in the cassette tray.
3)
Record the dimensions of the x-ray field
from the numerical indicators.
4)
Repeat steps l-3 with cassette rotated 90
degrees in the cassette tray.
5)
Repeat steps l-4 with a larger cassette, 30 x
35 cm (11 x 14") in the cassette tray.
9